2.2 Conductivity To TDS Conversion Factors

1. Factor

The conductivity to ppm TDS conversion factor. Multiply conductivity by this factor to get ppm TDS for the type of TDS reading needed. 

2. 442

A formulation that most closely represents the conductivity to ppm relationship, on average, for naturally occurring fresh water. 

Factor = Actual TDS / Actual Conductivity @ 25°C.

TABLE HERE

*Information Provided Courtesy of Eutech Instruments Pte Ltd Copyright © 1997

 

Conductivity to TDS Conversion Factors

Conductivity at 25°C TDS KCl   TDS NaCl   TDS 442  
  ppm Value Factor ppm Value Factor ppm Value Factor
84  µS 40.38 0.5048 38.04 0.4755 50.50 0.6563
447  µS 225.6 0.5047 215.5 0.4822 300.0 0.6712
1413  µS 744.7 0.5270 702.1 0.4969 1000 0.7078
1500  µS 757.1 0.5047 737.1 0.4914 1050 0.7000
8974  µS 5101 0.5685 4487 0.5000 7608 0.8478
12,880  µS 7447 0.5782 7230 0.5613 11,367 0.8825
15,000  µS 8759 0.5839 8532 0.5688 13,455 0.8970
80 mS 52,168 0.6521 48,384 0.6048 79,688 0.9961